CLIA test report requirements: the 7 elements every report must show
Every patient report issued by a laboratory holding a CLIA certificate must carry seven specific pieces of information, set out in 42 CFR 493.1291(c). They are short, and easy to miss when a report template is built or changed. Below is each element in plain English, the reporting duties that sit around it, and a checklist you can tick off against your own report and print for your quality file.
Reviewed 2026-11-08 against the sources listed at the end. General information for US laboratories, not legal advice.
The rule, in one paragraph
Under 42 CFR 493.1291 ("Standard: Test report"), "the test report must indicate" seven things: positive patient identification; the name and address of the laboratory location where the test was performed; the test report date; the test performed; the specimen source, when appropriate; the test result with units of measurement or interpretation where applicable; and any information on the condition and disposition of specimens that do not meet the laboratory's acceptability criteria. The same section also governs reference intervals, critical-value calls, delayed results, referral testing, corrected reports and who may receive results.
CLIA certificates are issued to laboratories, not to software. No LIS is "CLIA approved" — what software can do is make every report carry the required elements, every time, for every operator.
The seven required elements
| Citation | Element | What it means in practice |
|---|---|---|
| 493.1291(c)(1) | Positive patient identification | Either the patient's name and identification number, or a unique patient identifier and identification number. |
| 493.1291(c)(2) | Name and address of the testing location | "The name and address of the laboratory location where the test was performed" — the site that ran the test, which matters if you have more than one. |
| 493.1291(c)(3) | Test report date | The date of the report. |
| 493.1291(c)(4) | Test performed | The name of the test that was performed. |
| 493.1291(c)(5) | Specimen source, when appropriate | For example urine, wound swab or CSF — wherever the source affects how the result is read. |
| 493.1291(c)(6) | Result, with units or interpretation | The test result and, if applicable, the units of measurement or interpretation, or both. |
| 493.1291(c)(7) | Condition of unacceptable specimens | Any information about the condition and disposition of specimens that do not meet the lab's acceptability criteria (for example "hemolyzed — potassium not reported"). |
The reporting duties around the report
| Citation | Duty | What it means in practice |
|---|---|---|
| 493.1291(d) | Reference intervals available | Pertinent reference intervals ("normal" values) set by the performing lab must be available to the person who ordered the test. Most labs print them on the report. |
| 493.1291(g) | Critical values called immediately | Imminently life-threatening results, panic or alert values must be communicated to the ordering party immediately — and documented. |
| 493.1291(h) | Delayed results assessed | When results cannot be reported within your established time frames, decide by urgency whether to notify the ordering party of the delay. |
| 493.1291(i)(3) | Referral lab named | If you send specimens out, the ordering party must be told the name and address of each laboratory location that performed the test. |
| 493.1291(k) | Corrected reports keep the original | When a reported result is wrong: notify promptly, issue a corrected report promptly, and keep duplicates of both the original and the corrected report. |
| 493.1105(a)(6) | Reports retained | Keep, or be able to retrieve, a copy of each original, preliminary and corrected report for at least 2 years after reporting (10 years for pathology reports). |
Results may be released only to authorized persons — and, where applicable, to the person responsible for using them and the lab that referred the test (493.1291(f)); patients or their representatives may be given completed reports on request (493.1291(l)). The retention periods come from 42 CFR 493.1105.
Where report templates usually fall short
- A multi-site lab prints one head-office address on every report, instead of the location that actually performed the test.
- Specimen source is missing on microbiology and body-fluid tests, where it changes the interpretation.
- Rejected or compromised specimens (hemolyzed, clotted, insufficient) are simply left off the report rather than stated.
- A corrected report overwrites the original, so there is no copy of what the clinician first received.
- Send-out results are transcribed without naming the performing laboratory.
- Qualitative results print with no interpretation, or numeric ones without units.
What SamLab™ reports include
We checked the sample reports rendered by the SamLab™ app (see report formats) against the seven elements. The table shows what is printed out of the box and what you should configure in your own template. Your lab director remains responsible for the report your lab issues.
| Element | SamLab™ report |
|---|---|
| 493.1291(c)(1) — Positive patient identification | Printed: Patient name and patient ID are printed in the patient panel. |
| 493.1291(c)(2) — Name and address of the testing location | Printed: The lab name and address in the report header come from your lab profile — set them to the location that performed the test. |
| 493.1291(c)(3) — Test report date | Printed: The report date and time is printed ("Report Printed On"). |
| 493.1291(c)(4) — Test performed | Printed: Each test is printed by name above its results. |
| 493.1291(c)(5) — Specimen source, when appropriate | Configure: Configure your report template to show the specimen source wherever it matters for interpretation. |
| 493.1291(c)(6) — Result, with units or interpretation | Printed: Result, unit and reference interval are printed side by side. |
| 493.1291(c)(7) — Condition of unacceptable specimens | Configure: Configure your report template or result comments to state the condition and disposition of any specimen that fails your acceptability criteria. |
A verified result in SamLab™ locks, and a correction is saved as a new version with the original kept — which supports the record of original and corrected reports that 493.1291(k) asks for. More on records an inspector asks for: CLIA and CAP: what your records must show. For the full US overview, see SamLab™ for US laboratories.
Checklist: audit your report template
Take a recent report from each department and tick each item it shows. Nothing you tick is stored or sent.
0 of 13 checked
On every report — 42 CFR 493.1291(c)
Around the report
Report / department: ____________
Checked by: ____________
Date: ____________
Source: 42 CFR 493.1291 and 493.1105, read November 2026.
Questions labs ask
What must be on a CLIA lab report?+
42 CFR 493.1291(c) requires positive patient identification; the name and address of the testing location; the report date; the test performed; the specimen source when appropriate; the result with units or interpretation; and information on any specimen that failed acceptability criteria.
Do reference ranges have to be printed on the report?+
The rule (493.1291(d)) says pertinent reference intervals must be available to the person who ordered the test. Printing them on the report is the simplest way to meet it, and most labs do.
How long must a lab keep test reports under CLIA?+
42 CFR 493.1105(a)(6) requires keeping, or being able to retrieve, a copy of each original, preliminary and corrected report for at least 2 years after reporting, and pathology reports for at least 10 years.
Can lab software hold a CLIA certificate?+
No. CLIA certificates are issued to laboratories, not to software. An LIS can help by printing every required element consistently and keeping corrected reports alongside the originals.
Sources